The EDT directION Flow plus range incorporates two solid state ISE’s in one 12 mm Epoxy body. One acts as the Sensor whilst the other is incorporated into the reference system. Watch Set up Video
Flow plus combination ISE's do not require a separate reference electrode, making it convenient to use with small sample volumes. The sensor and reference are solid state which ensures a longer shelf life than any other type of ISE.
The reference half-cell has a free-flowing liquid junction which will reduce drift and provide more stable readings than conventional ISE’s. Flow plus electrodes are ideal for measuring awkward or dirty samples, which would clog conventional reference electrodes. Flow plus ISEs require a filling solution. The capacity of the reference chamber is 3ml.
The Flow plus reference system has a liquid junction provided by a conical plug which is opened or closed by rotation of the electrode cap. The wider the opening between body and plug the faster the electrolyte flows. Low ionic strength samples and samples with high viscosity or particulates are examples of media that require this increased flow.
Flow plus ISEs are ideal in situations where drift, slow response times and reduced electrode lifetimes are a problem. Each EDT directION Flow plus ISE has a 1 metre cable and a BNC connector. This allows for use with all types of pH/ION meter including laboratory bench and research models. Other connectors available on request.
For direct concentration reading and absolute simplicity the EDT QP459 pH/ION meter is the perfect instrument for these ISEs.
More Info:
EDT are a UK based manufacturer and produce ISEs for many prominent companies in the water testing sector. As a result our ISEs are very competitively priced and come complete with technical support provided from our vast library of applications and our in house electrochemistry technicians.
Use our ION Concentration Calculator for Quick Ion analysis for under £500.
Use EDT direction standards and Ionic Strength Adjustment Buffers ( ISABs) for optimum results. Click below to see a simple calibration procedure or learn more about the principles of ISE measurement.
Additional information
Weight | 0.15 kg |
---|---|
Dimensions | 26 × 9 × 4 cm |
Length | 155mm |
Diameter | 12mm |
Cap Diameter | 16mm |
Connector | BNC |
Cable Length | 1m |
Resistance at 25 Deg C | < 2.5 MOhm |
Concentration Range | 0.02-4010 ppm |
Potential Drift | 2mV per Day |
pH Range | 3.5-11 |
Temperature Range | 5 – 50°C |
Endpoint Time | Typically 10-60 seconds |
Interferences | Aluminium, Barium, Strontium |
Reference Type | Liquid Junction |
Commodity Code | 90279050 |