Potassium in Soil
Equipment Required
2. 3031 Potassium Combination ISE
3. 21314 Potassium 1000ppm Standard Solution
4. 30314 Potassium Ionic Strength Adjustment Buffer
5. Glassware. Beakers 250ml, Volumetric Flask 100ml. 10ml pipette
6. Deionised water.
Standard Preparation
Prepare standards of 10 and 100ppm Potassium by serial dilution of the 1000ppm standard. This is achieved by pipetting 10ml of the standard into a 100ml Volumetric flask and diluting to the mark with deionized water.
This is now a 100ppm Standard solution. Prepare further standards as required.
Sample preparation:
1. Air dry soil samples, crush and pass through a 2mm sieve.
2. Add 40ml of distilled water to a 20g soil sample.
3. Shake for 30 sec, stand for 15 mins and shake again for 5 sec. Repeat this shaking procedure three times. Allow soil to settle.
4. Extract 25ml of the supernatant sample from the top of the beaker.
This is your sample.
Method
To 25ml of each of the standards and samples add 1ml of 30314 ISAB and mix the contents. Be sure the beakers are clean and make sure not to touch the inside of the beakers with bare hands as contamination from sweat etc. is common. Beakers that have been washed with softened water or tap water will be contaminated. In these cases rinse the beakers with deionised water.
Select Potassium in the Ion mode of the QP459 Meter. Immerse the electrode in each of the standards in increasing concentration steps in the calibration mode rinsing the electrodes with distilled water and dabbing off the excess water between standards.
Using the QP459 Ion meter will allow the result to be read directly on the display. It also allows for up to a 3 point calibration if required.
Calculation:
Multiply the concentration on the QP459 display by 2 to allow for dilution.